_meta:
  source: OIML D 11:2013 "General requirements for electronic measuring instruments"
  source_path: mn-samples-oiml/sources/d011-e13/sections/
  description: |
    Environmental classification systems and performance tests defined by
    OIML D 11:2013, referenced by multiple Recommendations
    (R21, R46, R49, R50, R51, R60, R61, R76, R85, R99, R106, R107, R117,
    R129, R137, ...).

    Two layers of content:
      1. Environmental class systems — three orthogonal enums (H/M/E) used by
         Recommendations to declare the conditions an instrument is rated for.
      2. Performance tests — D 11 §9-§14, tables 6-41. Recommendations may
         cite any subset of these tests as having been applied during type
         evaluation, with a test_level index specifying severity.

    NOTE on R60 humidity marking: OIML R60 load cells use a SEPARATE
    humidity-marking system (CH/NH/SH) as part of the load cell classification
    symbol (R60:2021 §5.1.6, figure 4). CH/NH/SH is NOT the same concept as
    D 11 H1/H2/H3 — R60's marking is a property of the load cell per its
    classification symbol, while D 11 H-class is the instrument's installed
    location. CH/NH/SH stays in schema/R60.yaml.

environmental_classes:
  climatic_environment_class:
    description: |
      D 11 §7.2 table 1. Classification of ambient conditions related to
      humidity and water exposure at the instrument's installed location.
    enum: [H1, H2, H3]
    definitions:
      H1: Instruments used in temperature-controlled enclosed (weather protected)
        locations; local humidity not controlled; no exposure to condensed water,
        precipitation, or ice.
      H2: Instruments used in enclosed (weather protected) locations where local
        climate is not controlled; may be subject to condensed water, water from
        sources other than rain, and ice formations.
      H3: Instruments used in open air locations excluding extreme climate zones
        (polar/desert).
  mechanical_environment_class:
    description: |
      D 11 §7.3 table 2. Classification of ambient conditions related to
      vibration and shock at the instrument's installed location.
    enum: [M1, M2, M3]
    definitions:
      M1: Locations with vibration and shocks of low significance.
      M2: Locations with significant or high levels of vibration and shock
        (e.g. transmitted from machines and passing vehicles).
      M3: Locations where vibration and shock is high or very high (e.g.
        instruments directly mounted on machines, conveyor belts).
  electromagnetic_environment_class:
    description: |
      D 11 §7.4 table 3. Classification of expected electromagnetic
      environmental conditions.
    enum: [E1, E2, E3]
    definitions:
      E1: Locations where EM disturbances correspond to residential, commercial,
        and light industrial environments.
      E2: Locations where EM disturbances correspond to industrial buildings.
      E3: Measuring instruments powered by the battery of a vehicle and exposed
        to EM disturbances in any environment not hazardous for the general public.

checking_facility_type:
  description: |
    D 11 §3.19 / §5.3.1. Type of checking facility on a measuring instrument
    for detecting significant faults.
  enum: [P, I, N]
  definitions:
    P: Type P checking facility (Permanent).
    I: Type I checking facility (Intermittent).
    N: Type N checking facility (Non-durable, no checking facility).

test_level_index:
  description: |
    D 11 §7. Each test in D 11 has a recommended severity level indexed 1-5.
    A Recommendation declares which level applies for each test it cites.
  enum: [1, 2, 3, 4, 5]

tests:
  description: |
    Performance tests defined in D 11 §9-§14. Each entry is a single test
    identified by its D 11 table number. A Recommendation schema references
    a subset of these via `d011_tests:` and records the `test_level_index`
    actually applied during type evaluation.
  climatic:
    description: D 11 §9 — Climatic tests
    entries:
      d011_t06_static_temperature:
        table: 6
        name: Static temperature (low)
        reference_section: D 11 §9.1
      d011_t07_static_temperature:
        table: 7
        name: Static temperature (high)
        reference_section: D 11 §9.1
      d011_t08_damp_heat_steady_state:
        table: 8
        name: Damp heat, steady state
        reference_section: D 11 §9.2
      d011_t09_damp_heat_cyclic:
        table: 9
        name: Damp heat, cyclic
        reference_section: D 11 §9.2
      d011_t10_water:
        table: 10
        name: Water (rain)
        reference_section: D 11 §9.3
      d011_t11_atmospheric_pressure:
        table: 11
        name: Atmospheric pressure variation
        reference_section: D 11 §9.4
      d011_t12_vacuum:
        table: 12
        name: Vacuum
        reference_section: D 11 §9.4
      d011_t13_sand_dust:
        table: 13
        name: Sand and dust
        reference_section: D 11 §9.5
      d011_t14_salt_mist:
        table: 14
        name: Salt mist
        reference_section: D 11 §9.6
  mechanical:
    description: D 11 §10 — Mechanical tests
    entries:
      d011_t15_vibration_random:
        table: 15
        name: Vibration, random
        reference_section: D 11 §10.1
      d011_t16_vibration_sinusoidal:
        table: 16
        name: Vibration, sinusoidal
        reference_section: D 11 §10.1
      d011_t17_mechanical_shock:
        table: 17
        name: Mechanical shock
        reference_section: D 11 §10.2
  conducted_electromagnetic:
    description: D 11 §11 — Disturbances introduced through external wiring
    entries:
      d011_t18_dc_mains_voltage_variation:
        table: 18
        name: DC mains voltage variation
        reference_section: D 11 §11.1
      d011_t19_ripple_dc_mains:
        table: 19
        name: Ripple on DC mains power
        reference_section: D 11 §11.1
      d011_t20_ac_mains_voltage_variation:
        table: 20
        name: AC mains voltage variation
        reference_section: D 11 §11.2
      d011_t21_ac_mains_frequency_variation:
        table: 21
        name: AC mains frequency variation
        reference_section: D 11 §11.2
      d011_t22_dc_mains_dips_interruptions:
        table: 22
        name: DC mains voltage dips, short interruptions, voltage variations
        reference_section: D 11 §11.3
      d011_t23_ac_mains_dips_interruptions:
        table: 23
        name: AC mains voltage dips, short interruptions, reductions
        reference_section: D 11 §11.3
      d011_t24_ac_mains_harmonics:
        table: 24
        name: AC mains frequency harmonics
        reference_section: D 11 §11.3
      d011_t25_vlf_lf_disturbances_mains:
        table: 25
        name: VLF and LF disturbances on AC and DC mains
        reference_section: D 11 §11.3
      d011_t26_bursts_ac_dc_mains:
        table: 26
        name: Bursts (transients) on AC and DC mains
        reference_section: D 11 §11.3
      d011_t27_surges_ac_dc_mains:
        table: 27
        name: Surges on AC and DC mains power lines
        reference_section: D 11 §11.3
      d011_t28_bursts_signal_data_control:
        table: 28
        name: Bursts (transients) on signal, data and control lines
        reference_section: D 11 §11.4
      d011_t29_surges_signal_data_control:
        table: 29
        name: Surges on signal, data and control lines
        reference_section: D 11 §11.4
  radiated_electromagnetic:
    description: D 11 §12 — Electromagnetic fields
    entries:
      d011_t30_mains_power_frequency_em_field:
        table: 30
        name: AC mains power frequency electromagnetic field
        reference_section: D 11 §12.1
      d011_t31_conducted_rf_currents:
        table: 31
        name: Conducted (common mode) currents generated by RF EM fields
        reference_section: D 11 §12.2
      d011_t32_rf_em_fields_specific:
        table: 32
        name: RF EM fields (specific origin)
        reference_section: D 11 §12.2
      d011_t33_rf_em_fields_general:
        table: 33
        name: RF EM fields (general origin)
        reference_section: D 11 §12.2
      d011_t34_rf_em_fields_digital_radio:
        table: 34
        name: RF EM fields (digital radio telephones and portable radio transceivers)
        reference_section: D 11 §12.2
      d011_t35_electrostatic_discharges:
        table: 35
        name: Electrostatic discharges
        reference_section: D 11 §12.3
  battery:
    description: D 11 §13-§14 — Battery tests (general and road vehicle)
    entries:
      d011_t36_general_battery:
        table: 36
        name: General battery tests (disposable / rechargeable)
        reference_section: D 11 §13
      d011_t37_road_vehicle_battery_voltage_variation:
        table: 37
        name: Voltage variations of a road vehicle battery
        reference_section: D 11 §14.2
      d011_t38_road_vehicle_transient_supply_lines:
        table: 38
        name: Electrical transient conduction along supply lines (external 12V/24V)
        reference_section: D 11 §14.2
      d011_t39_road_vehicle_transient_other_lines:
        table: 39
        name: Electrical transient conduction via lines other than supply (12V/24V)
        reference_section: D 11 §14.2
      d011_t40_battery_voltage_cranking:
        table: 40
        name: Battery voltage variations during cranking
        reference_section: D 11 §14.2
      d011_t41_load_dump:
        table: 41
        name: Load dump test
        reference_section: D 11 §14.2

normalization_map:
  description: |
    Maps non-canonical surface forms seen in cert data to canonical D 11 enum
    values. Free-text humidity descriptions in older certs should be normalized
    via this map during migration.
  condensing: H2
  non-condensing: H1
  "non-condensing, indoor": H1
  "condensing, outdoor": H3

definitions:
  ClimaticClass:
    description: D 11 §7.2 table 1 — climatic environment class.
    allOf:
    - $ref: "_core.yaml#/definitions/StructuredValue"
    - properties:
        value:
          type: string
          oneOf:
          - const: H1
            description: Temperature-controlled enclosed (weather protected) locations
          - const: H2
            description: Enclosed (weather protected) locations without climate control; subject to condensation
          - const: H3
            description: Open-air locations excluding extreme climate zones (polar/desert)

  MechanicalClass:
    description: D 11 §7.3 table 2 — mechanical environment class.
    allOf:
    - $ref: "_core.yaml#/definitions/StructuredValue"
    - properties:
        value:
          type: string
          oneOf:
          - const: M1
            description: Locations with vibration and shocks of low significance
          - const: M2
            description: Significant or high levels of vibration and shock
          - const: M3
            description: High or very high vibration and shock (e.g. mounted on machines)

  EMClass:
    description: D 11 §7.4 table 3 — electromagnetic environment class.
    allOf:
    - $ref: "_core.yaml#/definitions/StructuredValue"
    - properties:
        value:
          type: string
          oneOf:
          - const: E1
            description: Residential, commercial, and light industrial environments
          - const: E2
            description: Industrial buildings
          - const: E3
            description: Vehicle battery-powered instruments in general-public environments

  TemperatureRange:
    description: Ambient operating temperature range, structured with unit.
    allOf:
    - $ref: "_core.yaml#/definitions/StructuredValue"
    - properties:
        unit:
          default:
            unit_id: u:degree_Celsius

  CheckingFacilityType:
    description: D 11 §3.19 / §5.3.1 — checking facility type.
    type: string
    oneOf:
    - const: P
      description: Type P — permanent checking facility
    - const: I
      description: Type I — intermittent checking facility
    - const: N
      description: Type N — non-durable / no checking facility

  TestLevelIndex:
    description: D 11 §7 test level index (1-5).
    type: integer
    minimum: 1
    maximum: 5

  TestReference:
    description: Reference to a D 11 performance test with optional level and notes.
    type: object
    required: [test_id]
    properties:
      test_id:          { type: string, pattern: '^d011_t\d{2}_' }
      test_level_index: { $ref: "#/definitions/TestLevelIndex" }
      notes:            { type: string }